Skip Navigation
Skip to contents

JYMS : Journal of Yeungnam Medical Science

Indexed in: ESCI, Scopus, PubMed,
PubMed Central, CAS, DOAJ, KCI
FREE article processing charge
OPEN ACCESS
SEARCH
Search

Articles

Page Path
HOME > J Yeungnam Med Sci > Volume 8(2); 1991 > Article
Original Article A study on dose distribution of small irradiation field in the electron therapy.
Sung Kyu Kim, Sei One Shin, Myung Se Kim
Journal of Yeungnam Medical Science 1991;8(2):114-120
DOI: https://doi.org/10.12701/yujm.1991.8.2.114
Published online: December 31, 1991
  • 1,396 Views
  • 1 Download
  • 0 Crossref
  • 0 Scopus

In electron therapy, low melting point alloy is used for shaping of the field. Electron field shaping material affect the output factor as well as the collimator system. The output factors of electron beams for shaped fields from NELAC-1018 were measured using ionization chamber of Farmer type in water phantom. The output factors of electron beams depend on the incident energy, inherent collimator system and the size of shaped field. Obtained results were followings. 1. In the smaller applicator, output varied extremely according to extent of collimator opening. 2. The higher energy, the output is less varied according to treatment field at small field.